Bayer HealthCare Rapidlab 800 User Manual

Page 110

Advertising
background image

2Ć12

Rapidlab 800 Operator's Manual

Figure 2Ć6.ā Sample Type Menu

Select Micro Sample.

3.

Select

Micro Sample

and press

Enter

.

NOTE:ĄĂ

If the sample device is a syringe, the system tries to move the sample

automatically for routine analysis. If the syringe contains sufficient volume,
the system analyzes the sample automatically.

4.

Press

Analyze

.

The Position Sample Manually message box appears.

5.

Turn the sample pump clockwise to move the leading edge of the sample to
point A, as shown in Figure 2-7.

Figure 2Ć7.ā Sample at Point A on an 850

Point A

Sample Pump

6.

Press

Sample in Place

.

7.

When prompted, remove the sample device.