Bayer HealthCare Rapidlab 800 User Manual

Page 457

Advertising
background image

System Administration

5Ć47

6.

Press

Yes

.

The QC File Setup screen appears with all data fields empty, as shown in
Figure 5-28.

Figure 5Ć28.ā QC File Setup (Ranges) Screen for an 860

Press Clear Entry to

delete text in a field.

7.

Complete the form.

If you use . . .

Then . . .

the keypad

type the low and high target limits for each parameter. Press

Enter

after you complete each field.

the optional bar
code scanner

scan the target range bar code for each parameter from QC
Expected Values. If you scan the wrong bar code, a message
appears prompting you to scan the correct parameter bar code.
Press

OK

and continue.

When you complete both fields for a parameter, the system automatically
calculates the target value and the action range.

8.

If a second screen of parameters exists, press

Next Screen

and enter values on

that screen.