Replacing the Sample Port

Materials required:

S sample port

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Pull off the sample door as shown in Figure 3-66.

Figure 3Ć66.āăReplacing the Sample Port

Drip Tray

Sample Door

Retainer Ring

Mount

Sample Port

OĆring

3.

Grasp the tab on the retainer ring and firmly pull the tab toward you to rotate
the ring.

4.

Grasp the sample port and attached drip tray and pull it off the mount as
shown in Figure 3-66.

NOTE:

Ensure that the three O-rings are in place.

5.

Install the new sample port on the mount.

6.

Push the tab on the retainer ring away from you until it locks into place.

7.

Reinstall the sample door, ensuring that it snaps in place.

8.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

9.

Press

Yes

to perform a two-point calibration.