800 Series Operator's Manual

5Ć50

Figure 5Ć29.ā QC File Setup Screen for an 850

This field displays

the file number that

you selected.

4.

Perform one of the following options.

If you want to . . .

Then . . .

change to a new
QC lot

type the new QC file information in the fields or, if you use
the optional bar code scanner, scan the new QC file
information. Continue with step 5.

edit the ranges for
this QC file

continue with step 5.

edit fields on this
screen

move to the field requiring a change and type the new data.
Press

Enter

after you complete each field. Continue with

step 5.

5.

Press

Done

.

A message appears prompting you to delete the current QC file setup data.

NOTE:ĄĂ

You must delete the current QC file data when you change lots.

Otherwise, the QC file will contain data from the old and the new lots.

If you are . . .

Then . . .

changing to a new
QC lot

press

Yes

.

The system prints a copy of the current QC file

setup data, deletes this data, and prints a copy of the QC
statistical summary report.

The QC File Setup (ranges) screen appears with all data fields
empty, as shown in Figure 5-30. Continue with step 6.

editing the existing
QC file ranges

press

No

. The QC File Setup (ranges) screen appears. The

fields contain the data for the file you selected. Continue with
step 6.