800 Series Operator's Manual

d. Take a reading of the actual sensor output signal.

e. Wait 100 seconds, and take another reading.

value = (reading 1 – reading 2) / 100 seconds

f. Calculate the resulting value and then compare the resulting value to the

expected value:

Sensor Circuit

Expected Value

pCO

2

–0.800 to +0.800 mV/s

pO

2

–0.300 to +0.300 nA

pH, Na

+

, K

+

, Ca

++

, or Cl

–0.545 to +0.545 mV/s

Glu I, Glu A

–5.348 to +5.348 nA

Lac I, Lac A

–5.348 to +5.348 nA

g. Press

Stop Test

.

CAUTION:

Touch the inner surface of the module frame to discharge static

buildup before removing or returning sensors.

6.

Reinstall the sensors:

a. Open the measurement module door.

b. Remove the test/blank sensor.

c. Reinstall the measurement sensor.

NOTE:ĄĂ

If you want to test another sensor, repeat steps 4b through 6c.

d. Reinstall the reference sensor if you removed it.

e. Close the measurement module door.

CAUTION:

Before you press

Exit Test

,

replace all sensors to prevent fluid from

spilling into the measurement module.

7.

Press

Exit Test

.

Perform a two-point calibration to ensure optimum performance.

NOTE:ĄĂ

If you encounter any difficulties when performing or interpreting the

results of this test, contact your Service Representative for assistance.

860

!

!