Testing a sensor circuit tools and supplies – Bayer HealthCare Rapidlab 800 User Manual

Page 358

Advertising
background image

4Ć72

800 Series Operator's Manual

Testing a Sensor Circuit

Tools and Supplies

S TB1, pO

2

/pCO

2

Test/Blank Sensor

S TB2, pH/Na

+

Test/Blank Sensor

S TB3, K

+

/Ca

++

/Cl

Test/Blank Sensor

S TB4, Glucose/Lactate Test/Blank Sensor
1.

Access the Measurement Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

3 Measurement

and press

Enter

.

CAUTION:

Touch the inner surface of the module frame to discharge static

buildup before removing or returning sensors.

2.

Remove the appropriate sensors:

a. Open the measurement module door.

b. If you are testing a pH or ISE sensor circuit, remove the reference sensor.

c. Remove the measurement sensor.

d. Close the measurement module door.

3.

Perform the Measurement Test without a sensor:

a. Press

Start Test

.

b. Take a reading of the actual sensor output signal.

c. Compare the result to the expected value.

NOTE:ĄĂ

Two values are reported for the glucose and lactate biosensors.

Sensor Circuit

Expected Value

pCO

2

–6.000 to +6.000 mV

pO

2

20.607 to 16.191 nA

pH, Na

+

, K

+

, Ca

++

, or Cl

–6.000 to +6.000 mV

Glu I

–5.348 to +5.348 nA

Glu A

25.252 to 54.748 nA

Lac I

–5.348 to +5.348 nA

(Continued)

3

Menu Code

3

!

860

860

860