Fluidics functions test – Bayer HealthCare Rapidlab 800 User Manual

Page 342

Advertising
background image

4Ć56

800 Series Operator's Manual

Test

Function

Roll Printer

Tests the ability of the roll printer to print all characters.

Bar Code Scanner

Tests the ability of the bar code scanner to read a test pattern.

Fluidics Functions Test

Use this procedure to test the ability of the reagent components to deliver reagents
to the fluid detectors in the reagent manifold and the measurement module.

1.

Access the Fluidics Functions Test screen from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

1 Fluidics Functions

and press

Enter

.

The Fluidics Functions Test screen appears as shown in Figure 4-2 for an 860
system.

Figure 4Ć2.ā Fluidics Functions Test Screen

Test one reagent

at a time.

Check here for results.

2.

Select the reagent you want to test and press

Enter

.

3.

Press

Start Test

.

4.

Check the screen for the message, Fluidics functions acceptable, to verify that
the system delivers the reagent to the fluid detectors.