Rapidlab 800 Operator's Manual

4.

Enter data in all required fields.

5.

Press

Done

.

The Patient Information screen closes.

If you want to make changes to the Patient Information screen during analysis,
press

Edit Patient Data

.

The Patient Information screen reappears. Make your

changes to the fields and press

Done

.

If you do not complete all required fields before you press Done, a message box
appears prompting you to complete the required data entry fields. Press

OK

to return

to the Patient Information screen. The incomplete field is highlighted.

If you do not complete data entry within 5 minutes, the system stores the sample
result, and enters a message containing a sequence number, the date, and the time
of analysis in the status log. The system returns to the Ready screen. You can
search for the sample results by sequence number and analysis time, and then enter
the patient data for the sample. Refer to Recalling Patient Sample Data, page 2-35.

You can enter a patient’s year of birth in a 2-digit or 4-digit format. However, the
system stores, prints, and transmits the birth year in a 4-digit format.

If you enter the birth year in a 2-digit format (xx), the system assumes the birth
year to be 19xx.

In the year 2000 and beyond, you must enter birth years in the 4-digit format to
ensure an accurate birthdate and patient age. The system verifies that the 4-digit
year entered is the current or a prior year. The valid range for years entered as 4
digits is 1880 to 2050.