Filling the measurement sensors – Bayer HealthCare Rapidlab 800 User Manual

Page 252

Advertising
background image

3Ć80

Rapidlab 800 Operator's Manual

Figure 3Ć55.āăFilling the KCl Reservoir

Electrode

Compartment

4M KCl Fill Solution

Fill Line

OĆring

OĆring

Sample Path

c. Tap the front face of the sensor with your knuckle to release any bubbles.

d. Gradually fill the KCl reservoir with KCl fill solution to the fill line.

CAUTION:ĄĂ

Do not overtighten the reservoir cap. Overtightening can damage

the gasket and cause leaks.

e. Reinstall the reservoir cap and hand-tighten.

11. Continue with

Reinstalling a Sensor

, page 3-89.

Filling the Measurement Sensors

Use this procedure to fill any measurement sensors that do not have sufficient fill
solution. Refer to Table 3-1 to determine the appropriate fill solution to use.

Materials required:

S appropriate fill solution
S lint-free tissue

NOTE:

If your system has a CO-ox module, follow the procedure described for

the appropriate base model. For example, information identified for an 860 also
applies to an 865.

NOTE:

The pO

2

and pCO

2

sensors do not require fill solution.

NOTE:

The glucose and lactate biosensors do not require fill solution.

!

860