Recalling qc data – Bayer HealthCare Rapidlab 800 User Manual

Page 143

Advertising
background image

2Ć45

Operating the System

Recalling QC Data

Use this procedure to recall the current month’s QC sample data and statistics.
When the system locates the QC sample data, you can edit the data, print reports,
and transmit the data to an LIS or data management system.

You can recall QC data for the previous month by archiving the previous month’s
QC data and then viewing the data. Refer to Archiving QC Data and Viewing
Archived QC Data
in Section 5.

1.

Access the Recall QC Data screen from the Menu screen:

2.

Select

4 Data Recall

and press

Enter

.

a. Select

2 QC Data

and press

Enter

.

The QC Data Search Criteria screen appears, as shown in Figure 2-25.

Figure 2Ć25.ā QC Data Search Criteria Screen

Type the number

to locate a

specific QC file.

This field displays the

number of QC samples

stored in the system.

Use the Date and Time fields

to recall reports for a shift,

day, week, or more.

NOTE:ĄĂ

The Analysis Date fields contain the current day’s date. If you do not

complete any other fields, the system recalls all QC samples for this day.

Use this screen to enter the criteria used to search the system for QC sample
data. The system uses all entered parameters when searching for patient data.
Samples meeting all criteria are recalled. For example, if you enter a QC file
number, the system searches for all samples stored in that QC file. If you enter
a file number and an analysis date, the system searches only for samples stored
in that QC file and analyzed on that date. If you do not complete any of the
criteria fields, the system recalls all the QC samples stored in the system.

4

Menu Code

2