Replacing the reference sensor cassette – Bayer HealthCare Rapidlab 800 User Manual

Page 246

Advertising
background image

3Ć74

Rapidlab 800 Operator's Manual

CAUTION:ĄĂ

Do not overtighten the reservoir cap. Overtightening can deform

the gasket and cause leaks.

e. Reinstall the reservoir cap and hand-tighten.

6.

Continue with

Reinstalling a Sensor

, page 3-89.

Replacing the Reference Sensor Cassette

Materials required:

S reference electrode refill kit
S hex tool
S lint-free tissue
1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

2.

Remove the reference sensor as shown in Figure 3-47:

a. Push up the latches on the measurement module door and lift the door.

b. Push the spring-loaded latch to the right.

c. Grasp the tab on the reference sensor and pull the sensor up and out of the

measurement module.

Figure 3Ć47.āăRemoving the Reference Sensor from an 850

Grasp the tab on the reference

sensor and pull it up and out of the

measurement module.

Push the springĆloaded latch

to the right.

!

2

Menu Code

7

!