Bayer HealthCare Rapidlab 800 User Manual

Page 404

Advertising
background image

4Ć116

800 Series Operator's Manual

Problem

Probable Cause and Solutions

If Blood, Question
Data

Optical measurements indicate that the CO-ox results should be
reviewed. A question mark (?) is printed next to the CO-ox
results on reports.

Results may still be clinically valid. When reporting CO-ox
results, consider the patient’s history and clinical condition to
determine acceptability of results.

To troubleshoot an If Blood, Question Data message,
determine whether the problem is with the sample type, the
patient sample, or the system.

NOTE:

When analyzed as a patient sample, aqueous materials

with dyes cause the If Blood, Question Data message to appear.

1.

Ensure that you analyzed a patient sample, not a QC or
calibration sample. Ensure that the patient sample is free of
clots and unidentified interfering substances.

2.

If the problem is not with the sample type, analyze the
sample again.

3.

If the message reappears, perform a pH/lytes one-point
calibration to zero the CO-oximeter

If the calibration . . .

is successful

is not successful

Then . .

go to step 5.

repeat the pH/lytes one-point
calibration two two more
times.

(Continued)