System Administration

5Ć19

Parameter

Action Range

Reference Range

Units

FO

2

Hb

%

94.0 – 97.0

[

–99.9 – 999.9

FCOHb

%

0.5 – 1.5

[

–99.9 – 999.9

FMetHb

%

0.0 – 1.5

[

–99.9 – 999.9

FHHb

%

0.0 – 5.0

–99.9 – 999.9

* Tietz NW ed. Fundamentals of clinical chemistry. 3rd ed. Philadelphia: WB Saunders, 1987; 864-891.

[

Weisberg HF. Acid-Base pathophysiology in the neonate and infant. Annals of Clinical and Laboratory
Science 1982; 12(4)249.

1.

Access the Reference and Action Ranges screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

2 Patient Data

and press

Enter

.

c. Select

1 Ref/Action Ranges

and press

Enter

.

The Reference and Action Ranges screen appears, as shown in Figure 5-9.

Figure 5Ć9.ā Reference and Action Ranges Screen for an 850

Press Clear Entry to

delete text in a field.

2.

Move to the field that you want to change.

3.

Type the new value and press

Enter

.

4.

Repeat steps 2 and 3 for any other values that you want to change.

5.

Press

Done

.

6.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

If you press Done before completing all required fields, the Incomplete Data Entry
message box appears. Press

OK

and complete the required fields.

If you type an invalid value in a field and press Done, the Invalid Range message
box appears. Press

OK

and type a valid entry in the field.

1

2

Menu Code

5