3Ć69

Maintaining the System

Filling the Reference Sensor

Materials required:

S reference electrode refill
S hex tool
S lint-free tissue

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

2.

Remove the reference sensor as shown in Figure 3-40:

a. Push up the latches on the measurement module door and lift the door.

b. Push the spring-loaded latch to the right.

c. Grasp the tab on the reference sensor and pull the sensor up and out of the

measurement module.

Figure 3Ć40.āăRemoving the Reference Sensor from an 850

Grasp the tab on the

reference sensor and

pull it up and out of the

measurement module.

Push the springĆloaded

latch to the right.

3.

Remove the reservoir cap from the reference sensor with the hex tool and set
the cap aside as shown in Figure 3-41.

2

Menu Code

7

!