3Ć83

Maintaining the System

Figure 3Ć58ā.ăFilling a Measurement Sensor

Fill Solution

OĆring

Electrode Compartment

7.

Insert the internal electrode into the electrode compartment and screw it into
place, ensuring that you do not cross-thread the electrode.

8.

Continue with

Reinstalling a Sensor

, page 3-89.

Replacing the Measurement or Sample Ground/

Temperature Sensors

If you need to replace the reference sensor, refer to

Replacing the Reference

Sensor

, page 3-71.

If you need to replace the glucose or lactate biosensors, refer to

Replacing the

Glucose and Lactate Biosensors

, page 3-86.

Materials required:

S appropriate sensor

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

NOTE:

If your system has a CO-ox module, follow the procedure described for

the appropriate base model. For example, information identified for an 860 also
applies to an 865.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

860

2

Menu Code

7