Valves test – Bayer HealthCare Rapidlab 800 User Manual

Page 348

Advertising
background image

4Ć62

800 Series Operator's Manual

11. Reinstall the reagent, sample, or waste pump tubing:

a. Disconnect the test tubing and the two-way connectors from the pump

tubing.

b. Reinstall the pump tubing:

S reagent pump tubing to the connectors at position 1
S sample pump tubing to the connectors at position 4
S waste pump tubing to the connectors at position 5

12. Reinstall the CO-ox tubing:

a. Disconnect the test tubing from the pump tubing.

b. Reconnect the sample tubing to the sample tubing connector on the pump.

c. Reconnect the waste tubing to the waste tubing connector on the pump.

13. Perform the appropriate action.

If you want to . . .

Then . . .

measure the flow rate
of a different pump

repeat steps 4 through 9.

return to the Menu
screen

a. Press

Exit Test

.

The Menu screen appears.

b. Press

Exit Menu

.

When you use the left arrow key to exit the Troubleshooting menu and enter the
Main menu, the system performs a wash.

Valves Test

Use this procedure to test the solenoid valve electronics and to test the ability of
the valves to open and close. You can test the following valves:

Valve

Function

6.8

controls the flow of 6.838 Buffer

7.3

controls the flow of 7.3/CO-ox Zero Buffer

Foam

controls the flow of Wash G/L Zero Reagent during the foam wash
sequence

Wash

controls the flow of Wash G/L Zero Reagent

(Continued)

Procedural

Notes

Advertising