D5 no endpoint, Problem, Solutions – Bayer HealthCare Rapidlab 800 User Manual

Page 308

Advertising
background image

4Ć24

800 Series Operator's Manual

D5 No Endpoint:

pO

2

pCO

2

Problem

Sensor does not reach stable reading within predefined time limit.

Qualifiers:

pO

2

pCO

2

Solutions

NOTE:ĄĂ

If D codes also exist for reagent problems such as D23, D24, or D29,

correct these problems first, then correct the D5.

If a D5 for pCO

2

occurs twice in a 4-hour period, perform QC analysis. If the

pCO

2

result is out of range, replace the sensor.

Ensure measurement module door was not opened during measurement. If door
was opened, repeat the sample analysis.

Perform a gas two-point calibration.

Deproteinize the sample path as described in Deproteinizing the Sample Path in
Section 3.

Perform the Measurement Test for the affected sensor as described in
Measurement Test, page 4-70.

Replace the affected sensor as described in Replacing the Measurement or
Sample Ground/ Temperature Sensors
in Section 3.

If the D code reappears, contact your Service Representative.

D5