Fluidic components, Sample chamber, Sample entry – Bayer HealthCare Rapidlab 800 User Manual

Page 40

Advertising
background image

1Ć18

Rapidlab 800 Operator's Manual

Sample Chamber

The hemoglobin content is measured in the sample chamber, which is located
between the fiber optics and the polychromator. After passing through the
hemolyzer, the sample moves to the sample chamber where it is warmed to 37

°

C.

Fluidic Components

The fluidic components move fluids, gases, and samples through the
800 system. The electronic components direct the fluidic components in response
to the user interface and automatic system activities. The CO-ox module uses the
fluidic components of the base model and additional components.

Sample Entry

The sample entry components, shown in the illustration below, include the sample
port and its components, the sample door, the sample probe, and the sample tubing.
You introduce samples into the system through these components.

Sample Door

Sample Probe

Sample Port

Sample Tubing

Retainer Ring

Sample Probe
Mount

Drip Tray

Capillary
Seal