Bayer HealthCare Rapidlab 800 User Manual

Page 564

Advertising
background image

EĆ38

800 Series Operator's Manual

Level

Recovery

Expected

Observed

WRSD

*

n

Mode

Capillary

23

0.44

50.3

49.9

100.8

Microcapillary

24

0.71

49.9

49.9

100.0

Microsyringe

24

0.54

49.8

49.9

99.7

Expired Gas

18

0.62

49.4

49.9

99.1

5

Syringe

24

0.52

71.3

71.3

100.1

Capillary

24

0.82

71.2

71.3

99.8

Microcapillary

24

1.07

71.0

71.3

99.5

Microsyringe

24

1.46

71.0

71.3

99.5

Expired Gas

24

0.24

70.5

71.3

98.8

* WRSD = within-run standard deviation

Table EĆ33.āă860 Recovery and Precision TestingĊpO

2

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

24

0.21

28.5

28.5

99.9

Capillary

24

0.25

28.5

28.5

100.1

Microcapillary

24

0.44

28.6

28.5

100.4

Microsyringe

24

0.33

28.7

28.5

100.7

Expired Gas

24

0.12

28.0

28.5

98.1

2

Syringe

24

0.35

50.2

49.9

100.6

Capillary

23

0.23

49.9

49.9

100.1

Microcapillary

24

0.38

50.0

49.9

100.2

Microsyringe

24

0.49

49.7

49.9

99.6

Expired Gas

24

0.56

49.8

49.9

99.7

(Continued)