3Ć107

Maintaining the System

12. Reconnect the power cord to the power source and allow the system to warm

up for at least 15 minutes.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

13. Press

No

.

14. Verify the system component temperatures from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

2 Temp/pAtm

and press

Enter

.

c. Press

Start Test

.

d. Verify that the system components warm up.

e. Press

Stop Test

.

f. If the temperature control system is off, press

Reset Control

and allow the

system to warm up.

g. Press

Exit Test

.

15. Perform a two-point calibration from the Menu screen:

a. Select

1 Calibration

, and press

Enter

.

b. Select

2 Two-point

, and press

Enter

.

3

Menu Code

2

1

Menu Code

2

(from the Main Menu)