Selecting parameter units – Bayer HealthCare Rapidlab 800 User Manual

Page 423

Advertising
background image

System Administration

5Ć13

Figure 5Ć5.ā Parameters Screen for an 850 system

Select the parameter

panel you want.

Press Next Screen

to view

additional parameters.

2.

Select the parameters.

a. Move to the frame that contains the appropriate parameter.

b. Select the parameter that you want and press

Enter

.

c. Repeat steps a and b for each parameter you want to change.

d. Press

Next Screen

to access the second screen of parameters.

e. Repeat steps a and b for each parameter you want to change.

f. Press

Next Screen

again to access the final screen of parameters.

g. Repeat steps a and b for each parameter you want to change.

3.

Press

Done

.

4.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

When you select a measured parameter to be on, you activate a sensor. Perform a
two-point calibration before you analyze a sample.

Selecting Parameter Units

Use this procedure to select the measurement units that the system displays for each
parameter. Table 5-4 lists the default and optional measurement units that you can
select for the base models and for the base models with a CO-ox module.

Table 5Ć4.āăParameter Measurement Units

Parameter

Default Units

Optional Units

System

pCO

2

, pO

2

mmHg

kPa

840, 850, 860

O

2

CT, BO

2

mL/dL

mL/L, mmol/L

840, 850, 860

(Continued)

Procedural

Notes

Advertising