Cleaning the reference sensor and removing bubbles – Bayer HealthCare Rapidlab 800 User Manual

Page 217

Advertising
background image

3Ć47

Maintaining the System

Cleaning the Reference Sensor and Removing

Bubbles

Use this procedure to clean the reference sensor and to remove bubbles at the
electrode tip and in the passage between the electrode compartment and the KCl
reservoir.

Materials required:

S reagent water
S lint-free swab

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Push up the latches on the measurement module door and lift the door.

3.

Push the spring-loaded latch to the right as shown in Figure 3-29.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

4.

Grasp the tab on the reference sensor and pull the sensor up and out of the
measurement module.

Figure 3Ć29.āăRemoving the Reference Sensor from the Measurement Module

Push the springĆloaded

latch to the right.

Grasp the tab on the reference

sensor and pull it up and out of the

measurement module.

5.

Check the sensor for bubbles at the electrode tip and between the electrode
compartment and the KCl reservoir as shown in Figure 3-30.