Analyzing qc samples – Bayer HealthCare Rapidlab 800 User Manual

Page 138

Advertising
background image

2Ć40

Rapidlab 800 Operator's Manual

Analyzing QC Samples

The 800 systems accept QC samples from a syringe or from an aspiration adapter.
When you perform routine QC analysis, all parameters available on your system
are analyzed. You can specify a panel of parameters to analyze for each sample.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Press

Change Sample Type

.

The Sample Type menu appears as shown in Figure 2-22.

Figure 2Ć22.ā Sample Type Menu

Select Quality Control (QC).

2.

Select

Quality Control (QC)

and press

Enter

.

NOTE:ĄĂ

If you are using Bayer Diagnostics controls, you can scan the bar code

on the ampule now.

If you want to . . .

Then . . .

analyze all parameters for
this sample

continue with step 3.

specify a panel of
parameters to analyze

a. Select

Change Parameters

. The Parameter Panel

appears.

b. Select the parameter panel and press

Enter

.

c. Continue with step 3.

3.

Prepare the sample and insert the device into the sample port. Figure 2-23
shows how to introduce a QC sample using an aspiration adapter.