Performance Characteristics

EĆ39

Level

Recovery

Expected

Observed

WRSD

*

n

Mode

3

Syringe

24

1.19

84.9

85.6

99.2

Capillary

24

0.61

85.4

85.6

99.7

Microcapillary

24

0.80

84.8

85.6

99.0

Microsyringe

24

1.08

85.2

85.6

99.5

Expired Gas

24

0.09

85.4

85.6

99.8

4

Syringe

24

1.34

150.0

149.7

100.2

Capillary

24

0.87

150.0

149.7

100.2

Microcapillary

24

2.04

150.1

149.7

100.2

Microsyringe

24

0.95

150.6

149.7

100.6

Expired Gas

24

0.19

150.0

149.7

100.2

5

Syringe

24

1.52

378.8

377.9

100.2

Capillary

24

1.55

378.7

377.9

100.2

Microcapillary

24

4.69

378.9

377.9

100.3

Microsyringe

24

2.35

379.6

377.9

100.5

Expired Gas

24

0.39

377.5

377.9

99.9

* WRSD = within-run standard deviation

Table EĆ34.āă860 Recovery and Precision TestingĊNa

+

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

119.4

Syringe

24

0.30

119.8

120.2

99.7

Capillary

24

1.33

120.0

119.8

100.2

Microcapillary

36

0.76

116.3

117.7

98.8

Microsyringe

24

0.47

119.4

120.3

99.2

(Continued)