Bayer HealthCare Rapidlab 800 User Manual

Page 244

Advertising
background image

3Ć72

Rapidlab 800 Operator's Manual

c. Slowly fill the electrode compartment by gently squeezing the container

until the KCl fill solution enters the KCl reservoir.

Figure 3Ć44.āăFilling the Electrode Compartment

OĆring

Sample Path

4M KCl Fill Solution

OĆring

Reference Sensor Cassette

(KCI Reservoir)

Electrode Compartment

CAUTION:ĄĂ

Do not touch the internal electrode wire. The wire is fragile and is

easily damaged.

4.

Install the internal electrode in the sensor:

a. Use the hex tool to remove the internal electrode from its container.

b. Insert the internal electrode into the electrode compartment.

c. Screw it into place with the hex tool, ensuring that you do not cross-thread

the electrode.

d. Tap the front face of the sensor with your knuckle to release any bubbles.

!