Replacing the measurement module lamp – Bayer HealthCare Rapidlab 800 User Manual

Page 274

Advertising
background image

3Ć102

Rapidlab 800 Operator's Manual

Replacing the Measurement Module Lamp

Materials required:

S measurement module lamp
1.

Shut down the system from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

3 Shutdown

and press

Enter

.

c. Press

Yes

.

CAUTION:

You must wait at least 1 minute before you disconnect the power cord,

and then wait at least 10 seconds before you reconnect the power cord. If you do
not adhere to the time intervals, you can damage the system.

2.

Wait until prompted, and then disconnect the power cord from the power
source.

3.

Disconnect the power cord from the side panel of the system.

4.

Turn the system around so that the rear panel faces you.

5.

Push the lamp cover up and remove the cover.

WARNING

Ensure that the lamp has been off for at least 5 minutes to allow

sufficient time for it to cool.

6.

Allow the lamp to cool for at least 5 minutes.

7.

Remove the connector from the system as shown in Figure 3-70.

Figure 3Ć70ā.ăReplacing the Measurement Module Lamp

Bracket

Lamp Cover

Connector

Clamp

Measurement

Module Lamp

7

Menu Code

3

!