2Ć32

Rapidlab 800 Operator's Manual

The following table describes the messages that may appear in patient sample
reports on the roll printer:

Message

Description

****** = Not in
Calibration

The sensor is out of calibration.

------

or ------

= Out of

Range: ____

The result is above or below the measurement range.

* = D5 No Endpoint: __

The sensor did not reach a stable reading within the
predefined time limit.

* = exception or D–code
noted (in Patient Data Log)

The sample has an exception or D code. Appears in
Patient Data Log.

or

= exceeds reference

range

The result is above or below the reference range.

↑↑

or

↓↓

= exceeds action

range

The result is above or below the action range.

? = If blood, question data

Optical measurements indicate that the CO-oximeter
results should be reviewed. See Troubleshooting Patient
Results
in Section 4.

SulfHb > 1.5%

The system detects a sulfhemoglobin concentration
greater than 1.5% in a CO-ox sample.

# = Interfering Substance:
Glu

The system detects the presence of substances in the
sample that may interfere with glucose measurement.

NOTE:

Repeated, unexpected occurrence of this

message may indicate sensor failure. See Troubleshooting
Patient Results
in Section 4.

# = Interfering Substance:
Lac

The system detects the presence of substances in the
sample that may interfere with lactate measurement.

NOTE:

Repeated, unexpected occurrence of this

message may indicate sensor failure. See Troubleshooting
Patient Results
in Section 4.

D38 Temperature Error

The system detects an error in the temperature control
system of the base model.

Bubbles Detected In
Sample

The system detects a non-continuous fluid in the sample
path of the base model.

(Continued)