Checking the Printer Paper

Check the amount of paper on the roll. A pink line on the paper indicates that the
roll is nearly empty. Replace the paper if the roll is empty or nearly empty as
described in

Replacing the Printer Paper

, page 3-66.

Checking the Waste Bottle Volume

Check the fluid volume in the waste bottle. Empty the waste bottle if it is at or near
the waste full line as described in

Emptying the Waste Bottle

, page 3-45.

Checking the Sample Path Temperature

1.

Access the Temperature screen from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

2 Temp/pAtm

and press

Enter

.

2.

Press

Start Test

.

3.

Press

Stop Test

.

4.

Check the screen to verify that the sample temperature is 37

±

0.15

°

C for the

base model and 37

±

0.35

°

C for the CO-ox module.

NOTE:

The 37

°

C temperature point is NIST traceable.

5.

Press

Exit Test

.

Checking the Barometer

Use this procedure to check the atmospheric pressure detected by the 800 system.

1.

Access the Barometer

Calibration screen from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

8 Barometer

and press

Enter

.