Bayer HealthCare Rapidlab 800 User Manual

Page 542

Advertising
background image

EĆ16

800 Series Operator's Manual

Level

Recovery

Expected

Observed

WRSD

*

n

Mode

4

Syringe

24

0.97

150.5

149.7

100.5

Capillary

18

0.86

150.2

149.7

100.3

Microcapillary

24

2.00

151.8

149.7

101.4

Microsyringe

24

0.80

150.6

149.7

100.6

Expired Gas

24

0.25

150.4

149.7

100.5

5

Syringe

24

1.20

377.9

377.9

100.0

Capillary

18

2.40

379.0

377.9

100.3

Microcapillary

24

4.08

380.7

377.9

100.7

Microsyringe

24

3.22

378.9

377.9

100.3

Expired Gas

24

0.66

378.7

377.9

100.2

* WRSD = within-run standard deviation