5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

984

■■■■■■■■

tiamo 2.2 (ProcessLab)

Identification

Description

.PK{#}.ABS

Absorbance of the peak with the greatest absorbance
(maximum absorbance) for evaluation # (1…9)

.PK{#}.TRN

Transmission of the peak with the greatest absorbance
(minimum transmission) for evaluation # (1…9) in %

.PK{#}.CNT

Intensity of the peak with the greatest absorbance
(maximum intensity) for evaluation # (1…9)

.PK{#}.CNR

Intensity in reference spectrum 'Command
name'.PK{#}.WVL for evaluation # (1…9)

.PK{#}.CND

Intensity in dark spectrum 'Command
name'.PK{#}.WVL for evaluation # (1…9)

.PK{#}.WVL

Wavelength of the peak with the greatest absorbance
for evaluation # (1...9) of the discovered maximum

.PK{#}.SAT

Specification as to whether the dectector was satura-
ted with 'Command name'.PK{#}.WVL for evaluation #
(1…9)

5.6.4.13.2

MEAS Spec - General/Hardware

Tab: Method

MEAS Spec Properties... General/Hardware

Command name

Name of the command.

Input

25 characters

The general parameters for the device are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start of
the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead