Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 847

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

831

Upper limit

Upper limit of the dosing rate. If the dosing rate exceeds this limit, then the
event Dosing rate upper limit exceeded will be triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

166.00 mL/min

Titrino
Range

0.00 ... 150.00 mL/min

Default value

150.00 mL/min

Upper hysteresis

Upper hysteresis of the dosing rate. When the dosing rate again falls below
the upper limit by this hysteresis value, then the event Dosing rate upper
limit ok
is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Action

Selection of the event to be triggered if the upper or lower limit is exceeded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running STAT command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running STAT command is quit, then the next command is executed.
Wait for [Continue]
Reagent dosing in the current STAT command is interrupted and a mes-
sage appears. As soon as the monitored measured value is again within
the limits (including hysteresis) reagent dosing can be resumed by press-
ing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.