Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 630

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

614

■■■■■■■■

tiamo 2.2 (ProcessLab)

EP recognition

Filter for the recognition of equivalence points:

Titrino
Selection

all | greatest | last | off

Default value

all

Titrando, 855
Selection

all | greatest | last | ascending | descending | off

Default value

all

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steepest
jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (Upol)

[New]

Opens the dialog window in which the parameters for a new window can
be entered (see Chapter 5.6.3.3.4.9, page 620).

[Properties]

Opens the dialog window Measured value # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.4.9, page 620).

[Delete]

Deletes the window selected in the table.

EP criterion

Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion will
not be recognized.

Range

1 ... 999

Default value

30

Advertising