5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

507

Stop measured value

Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Titrino
Range

-2000 ... 2000 mV

Selection

off

Default value

off

Stop EP

The titration is stopped when the specified number of equivalence points
has been found.

Range

1 ... 9

Default value

9

Selection

off

Volume after EP

When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.

Titrando, 855
Range

0.01000 ... 9999.99 mL

Selection

off

Default value

off

Stop time

Stops when the preset time (including start conditions) has elapsed since the
start of the titration.

Titrando, 855
Range

0 ... 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.