5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

954

■■■■■■■■

tiamo 2.2 (ProcessLab)

Command variables
The following command variables are generated in the method run by the
command MEAS Conc and can be used in formulas under the designation
'Command name.Variable designation':

Identifica-
tion

Description

.BLV

Blank value of the sensor used for the command (only for
ISE sensors)

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.DBL

Total duration for the processing of the command in s

.EME

Final measured value (measured value after processing of
the command) in the unit of the measured value

.ENP

Electrode slope of the sensor used for the command (in
mV for ISE sensors)

.ETE

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

.IME

Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value × (1...3) for the last measuring point of
the measuring point list

.LP.EXx

External value × (1...3) for the last measuring point of the
measuring point list

.LP.MEA

Measured value for the last measuring point of the meas-
uring point list in the unit of the measured value

.LP.TEM

Temperature for the last measuring point of the measur-
ing point list in °C

.LP.TIM

Time in s until the last measuring point of the measuring
point list is reached

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)