Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 360
4.6 Subwindow Information
■■■■■■■■■■■■■■■■■■■■■■
344
■■■■■■■■
tiamo 2.2 (ProcessLab)
Sensor 'Sensor name'
Information about the sensor used.
Sensor type:
Type of sensor (+ designation IS for intelligent sensors).
Ion:
Ion and valency for ISE electrodes.
Order number:
Order number of sensor.
Sensor serial number:
Serial number of sensor.
Device:
Name of device the sensor is connected to.
Measuring input:
Number of the measuring input to which the sensor is connected (+ desig-
nation iConnect for intelligent sensors).
Slope:
Electrode slope of the sensor used (in % for pH sensor or in mV for ISE).
E (0):
Electrode zero point of the sensor used (in mV only for ISE).
pH (0):
Electrode zero point of the sensor used (dimensionless for pH sensor).
c (blank):
Blank value of the ISE sensor used.
Cell constant:
Cell constant of the conductivity sensor used.
Calibration temperature:
Temperature during the calibration.
Calibration date:
Temperature during the calibration.