5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

867

Selection

'Sensor name' | not defined

Default value

not defined

not defined
If the sensor is not defined, then it will be checked at the start of the
determination whether a sensor of the type Thermoprobe is connected
to the measuring input. If this is the case, the connected sensor is used;
if it is no the case, an error message appears.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command.

5.6.3.9.3

TET - Start conditions

Tab: Method

TET Properties... Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the titration
is started.

Start volume

Start volume

Volume to be dosed with the indicated dosing rate before the start of the
titration .