5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

509

Evaluation without window

EP criterion

Criterion for the recognition of equivalence points. Equivalence points
whose ERC is smaller than the set EP criterion will not be recognized.

Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points:

Titrino
Selection

all | greatest | last | off

Default value

all

Titrando, 855
Selection

all | greatest | last | ascending | descending | off

Default value

all

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steepest
jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (Upol)

EP criterion

Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion will
not be recognized.

Titrino
Range

0 ... 200

Default value

5