5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

790

■■■■■■■■

tiamo 2.2 (ProcessLab)

Range

0 ... 99 s

Default value

0 s

Automatic start after sample addition

on | off (Default value: off)

If this option is enabled, the titration starts automatically when the command
is in status COND OK and the Threshold value set is exceeded. In this case
the system assumes that the sample has been injected. This parameter is
displayed only for 851 and 852.

Threshold value

Value the automatic start after the sample addition is carried out automat-
ically when it is exceeded.

851, 852
Range

0 ... 999 mV

Default value

50 mV

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the Event
message
Start titration is sent to the command with a SEND command.

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titration
will be ignored.

If this command is already started (e.g. via Automatic start after sam-
ple addition
), the SEND command will not have any effect.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.