6 meas t - fix endpoint evaluation, 8 meas t/flow, 1 meas t/flow - overview – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 963: Meas t/flow

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

947

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 880)
.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.4.7.6

MEAS T - Fix endpoint evaluation

Dialog window: Method

MEAS pH Properties... Additional evaluations Fix

endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-20.0 ... 150.0 °C

Time
Range

0.0 ... 999999.9 s

5.6.4.8

MEAS T/Flow

5.6.4.8.1

MEAS T/Flow - Overview

Dialog window: Method

MEAS T/Flow Properties... MEAS T/Flow - 'Command

name'

Command for Temperature and gas flow measurements.

Devices
This command can be executed with the following devices:

Sample Processor: 774, 874

Appearance
The command has the following appearance: