9.3 Determinations

■■■■■■■■■■■■■■■■■■■■■■

1566

■■■■■■■■

tiamo 2.2 (ProcessLab)

5 If a method was loaded in which the option Statistics in the START

command is enabled, disable the check box Statistics on the tab
Determination series.

6 Click on [Start].

The first determination of a sample series will be started. The sub-
window Method shows active tracks and commands, the subwindow
Live display live curves, measured values and messages.

Determination series with statistics

1 Select program part Workplace.

2 Use the Run subwindow to select the Determination series tab.

3 If desired, enter Determination parameters.

4 Load existing sample table or enter sample data directly into the work-

ing sample table.

The subwindow Method shows the method loaded in the first line.

Note

In order to carry out statistics calculations for results, the check box
Statistics has to be enabled in the START command of the
method. In addition, the check box Statistics has to be enabled for
every desired result in the CALC command.

5 On the Determination series tab, enable the check box Statistics.

6 If desired, modify number of single determinations.

7 Click on [Start].

The first determination of a sample series will be started. The sub-
window Method shows active tracks and commands, the subwindow
Live display live curves, measured values and messages.