3 kft ipol - start conditions, Start conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 759

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

743

the command. This parameter is displayed only for 751, 758, 784, 795 and
799.

5.6.3.5.2.3

KFT Ipol - Start conditions

Tab: Method

KFT Ipol Properties... Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the titration
is started.

Initial measured value

Note

Is displayed only for Titrandos.

The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:

Measured value drift

The measured value is only accepted if the drift is less than the value entered
here.

Range

0.1 ... 999.0 mV/min

Selection

off

Default value

off

off
The measured vales will not be applied until after the maximum waiting
time has passed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s