10 met ph - volume window – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 586

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

570

■■■■■■■■

tiamo 2.2 (ProcessLab)

ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.3.2.10

MET pH - Volume window

Dialog window: Method

MET pH Properties... Potentiometric evalua-

tion

Evaluation with volume window (mL) [New]/[Properties] Volume

window #

Volume windows are defined sub-regions (windows) on the volume axis for
which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.

Lower limit

Lower limit for the volume window.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Upper limit

Upper limit for the volume window.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

9999.99 mL

EP criterion pH

Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.

Titrando, 855
Range

0.1 ... 9.99

Default value

0.50

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first