5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1188

■■■■■■■■

tiamo 2.2 (ProcessLab)

The LOOP command is at the end of the loop for Repeat loops. As soon
as one stop criterion is fulfilled (case 2), the command after the LOOP com-
mand is carried out. If no stop criterion is fulfilled, (case 1), the loop is run
through again. In either case, the loop is run through at least once.

The LOOP command is at the beginning of the loop for While loops. If the
arrow has been dragged to a command after the LOOP command, another
arrow is automatically set from the previous command back to the LOOP
command. As soon as one stop criterion is fulfilled (case 2), the command
after the loop is carried out. If no stop criterion is fulfilled, (case 1), the loop
is run through. It is possible with certain stop criteria that the loop is never
run through.

Note

Nested and overlapping loops are not allowed.

Parameters
The parameters for the command LOOP are configured in the following
dialog window:

LOOP

Command variables
The following command variables are generated in the method run by the
command LOOP and can be used in formulas under the designation 'Com-
mand name.Variable designation'
:

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Command
has never been ended

.LCO

Loop counter = current quantity of completed runs, both of
Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

Advertising