4 meas opt - evaluations – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1007

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

991

5.6.4.14.4

MEAS Opt - Evaluations

Tab: Method

MEAS Opt Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves can
be activated and defined on this tab.

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (Measured
value
or Time) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fix
endpoint (see Chapter 5.6.4.14.5, page 993).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fix end-
point selected in the table (see Chapter 5.6.4.14.5, page 993).

[Delete]

Deletes the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the minimum measured
value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.