background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1085

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Upper limit

Upper limit of the temperature. When the temperature exceeds this value,
the action Temperature upper limit exceeded is triggered.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

150.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

500.0 °C

Upper hysteresis

Upper hysteresis of the temperature. When the temperature once again falls
below the upper limit by this hysteresis value, the action Temperature
upper limit ok
is triggered

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Action

Selection of the event to be triggered if the upper or lower limit is exceeded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is executed.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a mes-
sage appears. As soon as the monitored temperature is again within the
limits (including hysteresis) reagent dosing can be resumed by pressing
on [Continue] in this message box.

Advertising