Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 814

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

798

■■■■■■■■

tiamo 2.2 (ProcessLab)

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Cell

Generator type

Selection of the generator type.

Selection

without diaphragm | with diaphragm

Default value

with diaphragm

Generator current

Polarization current at the generator electrode.

Selection

100 mA | 200 mA | 400 mA | auto

Default value

auto

auto
Means that the current is automatically adapted to the conductivity of
the reagent and that in the region of the endpoint the current will be
controlled at smaller values

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Advertising