8 det ipol - additional measured values – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 530

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

514

■■■■■■■■

tiamo 2.2 (ProcessLab)

Lower limit

Value for the lower limit of the Gran evaluation range.

Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Upper limit

Value for the upper limit of the Gran evaluation range.

Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

5.6.3.2.4.8

DET Ipol - Additional measured values

Tab: Method

DET Ipol Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings:

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can be
calculated from existing measured values and saved, which can also be pre-
sented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.

[New]

Opens the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 879)
.

[Properties]

Opens the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 879)
.

[Delete]

Deletes the calculated measured value selected in the table.