5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1212
■■■■■■■■
tiamo 2.2 (ProcessLab)
Selection
s | min
Default value
s
5.6.11.7
PORT
5.6.11.7.1
PORT - Overview
Dialog window: Method
▶ PORT ▶ Properties... ▶ PORT - 'Command name'
Command for changing the port position on a multi-port valve.
Devices
This command can be executed with the following devices:
Titrando:901, 902, 904, 905, 906, 907
Appearance
The command has the following appearance:
Parameters
The parameters for the command PORT are configured in the following
dialog window:
■
Command variables
The following command variables are generated in the method run by the
command PORT and can be used in formulas under the designation 'Com-
mand name.Variable designation':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.EPP
Position after finishing the command: 0 = invalid position
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended