Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 784

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

768

■■■■■■■■

tiamo 2.2 (ProcessLab)

time
The endpoint must be exceeded for a certain time: the Delay time. The
same Delay time with different, very small volume increments (depending
on the volume of the exchange or dosing unit) means different switch-
off points.

Note

The Stop conditions will always stop the titration even if the stop criterion
has not been fulfilled or has been switched off.

Stop drift

Titration is stopped when the endpoint and the stop drift have been reached.
This parameter is only shown for Stop criterion = drift.

Range

1 ... 999 µL/min

Default value

20 µL/min

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Titrando
Range

1 ... 999 µL/min

Default value

10 µL/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is only
shown for Stop criterion = time. The selection inf. stands for infinite.

Titrando, 855
Range

0 ... 999 s

Default value

10 s

Titrino
Range

0 ... 999 s

Default value

10 s

Selection