5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

665

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Range

0.1 ... 999999 s

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s

Default value

2.00 s

5.6.3.4.3.6

SET pH - Stop conditions

Tab: Method

SET pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the titration are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Stops when the preset time (including start conditions) has elapsed since the
start of the titration.